Pacific Nanotechnology Inc.
 

Dual Scanner Technology in an AFM



Introduction

In roughly 20 years the atomic force microscopes moved from an esoteric invention to a routinely used tool for nanotechnology research and development. The AFM is credited with enabling many of the nanotechnology advancements in science and engineering. The AFM has applications in almost all science and engineering disciplines ranging from metrological measurements to visualizing nanostructures.

In the first AFM designs, derived from the scanning tunneling microscope, a light lever force sensor is used to measure the force between a probe at the end of a cantilever and a sample's surface. In these initial designs, the sample was scanned in the X, Y, and Z directions. Typically a piezoelectric tube was used for scanning the sample. Although fairly simple in design, this type of microscope proved the usefulness of the AFM for research and engineering. The tube scanner could be made relatively small and thus had a very high resonant frequency that facilitated fast scanning.

The greatest drawback of the sample scanning AFM using a piezo tube was the coupling that occurred between the X, Y and Z axis. Although the sample scanning AFM was very helpful for visualizing nanostructures, it was very difficult to make metrological measurements with an AFM having a tube as the primary scanner. A second drawback of sample scanning atomic force microscopes is that they are useful for only small samples.

Atomic force microscopes that scanned the probe while the sample remained stationary were first introduced almost 20 years ago. Probe scanning microscopes had the primary advantage that they could scan any sized samples. However, the structure of the tip scanning AFM is typically much larger than the sample scanning AFM. Thus, the resolution of the tip scanning AFM is often lower than the sample scanning AFM. Presently, most AFMs that are used for metrological applications scan the probe.

Design

Dual Scanner Technology (DST), innovated by Pacific Nanotechnology, gives researchers the full range of scanning options because the DST instrument incorporates both probe and sample scanning capabilities. Often a research project requires both capabilities in the same microscope. Figure 1 illustrates the design utilized in the Pacific Nanotechnology Nano-DSTT AFM. The sample is supported on a piezoelectric scanner, the rapid scanner. Above the sample is a probe scanning, light lever AFM.

Figure 2 illustrates the electronic control system for an AFM with Dual Scanner Technology. A control CPU is used to drive two independent XYZ boards. Each XYZ board has all of the electronic functions required to run an AFM including a PID control loop, phase/amplitude detection circuit, and XY high voltage circuits. The XYZ boards then drive the metrology and rapid scanners. Of course, specialized software is essential for operating an AFM having DST. The Control CPU must have a multi-threaded operating system such as LINUX so that each of the scanners can be driven independently.

A piezoelectric driven flexure metrological scanner is used for the metrological applications. Its XY scan range is typically 90 microns but can be as large as 360 microns. A range of 8 microns is available with the Z piezoelectric element. Calibration sensors in the X, Y and Z axes facilitate accurate metrological measurements and positioning of the AFM probe. The flexure scanner advantage over traditional piezoelectric tube scanners is that they have a minimal amount of bow in the Z axis. Additionally the flexure scanner has a minimal amount of coupling between the X, Y, and Z axis.

The rapid scanner has an XY range of 2 microns and a Z dynamic range of about 0.5 microns. Because the scanner has a limited dynamic range, it has a very high resolution in the XY axis and also a very low vertical noise floor. Additionally, the rapid scanner has a very high resonant frequency and can be scanned in the XY axis at rates as high 250 Hz. The Z feedback for the AFM can be derived from either the metrology or the rapid scanner.

Performance

A benchmark sample for showing the performance of an AFM sample is cleaved Mica. Viewing atomic corrugations of such a sample is only possible if the noise floor of the system is substantially below 0.08 nm and the horizontal resolution is substantially below 0.015 nm. Figure 3 illustrates an image of Mica acquired with the rapid scanner. The corrugations are clearly visible. This image was taken with both the metrological and the rapid scanner in the microscope. Although it is sometimes possible, it is very difficult to measure images of mica atomic corrugations with an AFM having a scan range in the XY axis that is greater than 50 microns. This is because of the dynamic range of the scanner.

There are several specifications that are critical for a high performance metrological AFM scanner including bow, linearity, and cross talk. There are samples available for demonstrating each of these specifications. For example, VLSI standard test patterns are used for demonstrating linearity and calibration in the XY and Z axis. Scanning a flat piece of silicon will demonstrate the residual bow in a scanner.

One of the most challenging samples to measure with an AFM metrological scanner is a micro-fabricated "triangle" sample. The sample is comprised of a series of parallel triangular patterns. The "triangle" sample will clearly illustrate the residual coupling between the XZ and YZ axis. Figure 4 is the line profile of a "triangle" sample measured with the flexure Nano-DSTT scanner.
The DST architecture further facilitates dynamic experiments. Such experiments take advantage of the relatively high speed scanning capabilities of the tube scanner in combination with the wide scanning range of the metrology scanner. Such measurements are made by scanning the rapid scanner at a rate of one frame per second and then using the metrology scanner to pan around the sample's surface. This technique is very helpful for scanning at high resolutions and finding nanostructures on a surface that are present in relatively low densities. Additionally, this method can be helpful for studying dynamic processes that occur on a sample surface.

Summary

Dual scanning technology is essential for researchers and engineers with the most demanding AFM applications. A DST atomic force microscope incorporates two scanners: a flexure scanner for precise metrology/positioning and a rapid scanner for the highest resolution scanning. Such as system is flexible so that any imaginable research project can be completed successfully without compromise.